1 Electrical and Computer Engineering Department, Johns Hopkins University, Baltimore, MD 21218, USA
2 Sheikh Zayed Institute for Pediatric Surgical Innovation, Children’s National Health System, Washington, DC 20010, USA
An endoscopic imaging system using a plenoptic technique to reconstruct 3-D information is demonstrated and analyzed in this Letter. The proposed setup integrates a clinical surgical endoscope with a plenoptic camera to achieve a depth accuracy error of about 1 mm and a precision error of about 2 mm, within a 25 mm×25 mm field of view, operating at 11 frames per second.
170.2150 Endoscopic imaging 170.4580 Optical diagnostics for medicine 110.6880 Three-dimensional image acquisition 120.3890 Medical optics instrumentation 